Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

BioMed has asked us for specifications for the Air Supply. What are they?

0
Posted

BioMed has asked us for specifications for the Air Supply. What are they?

0

The Specifications for the HoverMatt Air Supply are: • The hose end temperature should not exceed 110 degrees Fahrenheit after ten (10) minutes of running with no back pressure on the motor. • The hose end air pressure should be approximately 2.8 psi or higher and the volume should be approximately 78 cfm or higher.