Can 0.2 ml tubes be used in the Concentrator plus?
Yes, one can centrifuge 0.2 ml tubes in the 1.5 ml rotor F-45-48-11 with the help of an adapter #5425 715.005 (order no North America #022636260), in the 0.5 ml rotor F-45-72-8 with the help of the adapter #5425 723.008 (order no North America #022636286) or in the swing bucket rotor A-2-VC with the help of the work tray for 96 x 0.2 ml PCR tubes (#0030 124.235, order no North America #951010031).