Does the abnormal signal-averaged electrocardiogram predict future appropriate therapy in patients with implantable cardioverter-defibrillators?
GROUND: Several studies have documented the prognostic significance of the signal-averaged electrocardiogram (SAECG) both after myocardial infarction and nonischemic cardiomyopathy. However, whether the SAECG can identify patients with implantable cardioverter-defibrillator (ICD) who receive appropriate therapy has not been hitherto completely investigated. METHODS: Between August 2002 and August 2004, 83 consecutive ICD patients who had had SAECGs recorded were enrolled in this study. All patients were followed up in the outpatient ICD clinic, and interrogated electrograms were collected. RESULTS: Over 9.0 +/- 2.8 months of follow-up, 27 (32%) patients had appropriate ICD therapy for ventricular tachycardia or fibrillation; 15 (55.6%) patients had abnormal; and the remaining 12 (44.4%) had normal SAECGs. Of the 56 patients with no appropriate therapy, 27 (48.2%) and 29 (51.8%) patients had abnormal and normal SAECGs, respectively. There were no statistically significant differences be