Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Does the length of plastic pressure tubing between the pressure sensor being tested and the calibrator affect the operation and accuracy of the 869?

0
Posted

Does the length of plastic pressure tubing between the pressure sensor being tested and the calibrator affect the operation and accuracy of the 869?

0

Longer tubing increases the pump-up volume that the calibrator needs to drive in order to reach test pressure points. Normally, this would significantly increase the time to achieve set point pressures and the entire calibration test cycle. However, the 869 has a very fast micro-stepping motor driven piston/cylinder and is not appreciably slowed down by tubing lengths in excess of 12. The other concern with using long runs of tubing is that tubing movement can actually produce incremental pressure fluctuations. To reduce tubing pressure fluctuations, use the shortest length tubing possible or tie down longer runs to minimize tubing movement. The Setra calibrator will provide accurate sampling by filtering out pressure fluctuations.