Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Does the OD 600 method take place linearly on the BioPhotometer?

0
Posted

Does the OD 600 method take place linearly on the BioPhotometer?

0

This measuring method is principally non-linear, when one observes the absorption curve over wide OD ranges and dilutions. The values are dependent on both the organism (size, shape) and the device (light path geometry, ratio of measured stray light radiation). A correct procedure includes the generation of a calibration graph for each organism and device by determining the actual number of cells under a microscope. When the curve shape is known it is then possible to measure in ranges that are almost linear. Applications on the topic of the dependency of OD values on the organism and device are listed on www.biophotometer.info. Just click on “More information”.