Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

During my blank measurement on the BioPhotometer only “++++” appears on the display instead of 0.000 A. What happened?

0
Posted

During my blank measurement on the BioPhotometer only “++++” appears on the display instead of 0.000 A. What happened?

0

This is caused by the measurement outside of the upper measuring range (> 3.0 A). Please check 1. Whether your cuvette is pressed firmly onto the base of the cuvette shaft. The error message “++++” indicates that the cuvette base is blocking the light beam. 2. Whether your cuvette has an optical window at 8.5 mm. 3. Whether you have sufficient blank medium in the cuvette.