Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

For the main unit: How much lag time is there when inputting or outputting a signal from the external terminal of the measurement station?

0
Posted

For the main unit: How much lag time is there when inputting or outputting a signal from the external terminal of the measurement station?

0

The lag time from the external terminal to the bus trigger / time base is as follows: When inputting from TRIG IN: TRIG IN -> bus trigger signal : 40 ns (typical value) TRIG IN -> time base signal : 25 ns (typical value) When inputting from EXT. I/O: EXT. I/O -> bus trigger signal : 25 ns (typical value) EXT. I/O ->time base signal : 10ns (typical value) When outputting to EXT. I/O: EXT. I/O <- bus trigger signal : 25ns (typical value) EXT. I/O <- time base signal : 25ns (typical value) The lag time between the bus trigger / time base and each measurement module differs depending on the module. See the following example: Bus trigger signal -> WE7111 Digital oscilloscope module : 30 ns + sampling interval (typical value*) * Typical value represents a typical or average value. It is not strictly guaranteed.