Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How do you determine if rod lenses and/or fibers are excessively discolored? What tests are conducted to confirm fitness for use of these components?

0
Posted

How do you determine if rod lenses and/or fibers are excessively discolored? What tests are conducted to confirm fitness for use of these components?

0

The proper transmission of color for rod lenses and fibers is evaluated using visual control and photo analysis. Specifically, the customized IMS target contains different circles representing a color spectrum. Technicians visually compare the actual target colors to the colors as seen through the endoscope. If needed, a photograph of the target as seen through the endoscope is made for comparison to the actual target colors. Any unacceptable discoloration is noted via this visual test.