Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

I would like to perform precision measurement with my BioPhotometer using the “Function” menu. What should I bear in mind here and which solutions are suitable for such measurement?

0
Posted

I would like to perform precision measurement with my BioPhotometer using the “Function” menu. What should I bear in mind here and which solutions are suitable for such measurement?

0

Precision measurement should be carried out using a semimicro-quartz glass cuvette (> 350 µl). Clear and clean solutions are suitable as the measuring medium, for example potassium dichromate (zero adjustment to water) or water (zero adjustment to air). The coefficient of variation (CV) should be smaller/equal to 1% at 1 A in line with the technical data for the BioPhotometer. Please bear in mind that the Secondary UV-VIS filter should be used for exact precision measurement and determination of both the photometric and wavelength accuracy.