Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

On a network analyzer, what happens to the two-port calibration if the source or receiver attenuation is changed?

0
10 Posted

On a network analyzer, what happens to the two-port calibration if the source or receiver attenuation is changed?

0
10

The calibration is invalidated if you change the attenuator settings. If you change attenuator settings after you have performed a calibration, you must perform another calibration.