Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

The SCD/NCD technology is equimolar. Are there situations where an equimolar response will not be achieved?

0
Posted

The SCD/NCD technology is equimolar. Are there situations where an equimolar response will not be achieved?

0

Yes. The equimolar response will only be achieved for those analytes that are delivered to the burner without loss in the injection port, column, etc. Therefore, inactive or treated surfaces should be used. This includes connective tubing between valves, injection port liners, etc.