Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What can cause a sudden increase or decrease in O2 reading in the 3010MA?

cause increase O2 Reading Sudden
0
Posted

What can cause a sudden increase or decrease in O2 reading in the 3010MA?

0

Factors that can cause this are: — Mechanical vibration or shock — A sudden change is sample pressure, vent pressure, or flow rate — Particulate contamination of some type — Large fluctuation of power supply voltage Top of page