Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What happens if the air compressor malfunctions or stops working?If a required air dryer fails, damage to the PSA oxygen generator can occur due to moisture contamination of the sieve. Do you size systems differently for different altitudes?

0
Posted

What happens if the air compressor malfunctions or stops working?If a required air dryer fails, damage to the PSA oxygen generator can occur due to moisture contamination of the sieve. Do you size systems differently for different altitudes?

0

The size of the air compressor for a PSA oxygen generator has to be factored for site elevation. Higher elevations will require a larger air compressor.