Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is an appropriate set of 4-bar targets for MRTD (Minimum Resolvable Temperature Difference) measurement?

0
Posted

What is an appropriate set of 4-bar targets for MRTD (Minimum Resolvable Temperature Difference) measurement?

0

There are two references for selecting a set of 4 bar targets with the spatial frequencies covering a customer s range of interest. The Tri-Services Working Group recommends 4 spatial frequencies in their Guide for Preparing Specifications for Thermal Imaging Sets. These are recommended frequencies, not requirements. They are: 0.2 fo 0.5 fo 1.0 fo 1.2 fo fo is 1/(2 IFOV) where IFOV is the Instantaneous Field of View, (also called the DAS, Detector Angular Subtence). Holst simply refers to measuring MRTD in 3 places. A high frequency measurement (the value is not specified), a mid frequency test (the fo value described above), and a low frequency measurement which he has dependent on viewing distance. Since fo is the theoretical Nyquist frequency limit if a starring array has a 100% fill factor, the 1.0fo and 1.2fo values are adequate for the high frequency limit. The 1.2fo value is somewhat arbitrary, so a range of 1.2fo to 1.5fo should be acceptable (and is actually discussed in the T

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.