What is minimum weight moisture content rate measured at?
• In the case of using MX/MF-50, sample more than 0.1g is measurable. Sample mark on LCD display will light when sample amount is enough. Back to top • What is maximum weight moisture content measured at? • In the case of using MX/MF-50, sample less than 50g is measurable. “E” will be displayed if sample is over 50g. Back to top • Is it true that the more sample the more accurate the measurement will be? • No, that’s not true. Excess of a sample may not get heated inside evenly or heating time may be long. In such cases measurement may be not with high repeatability. Back to top • Does how to place sample such as powder on a pan affect measurement result? • Yes. How to put a sample on a pan determines heat distribution. To measure moisture content rate accurately even heating and vaporization are needed so sample should be flat on a pan.
Related Questions
- Why is the moisture content rate MS/MX/MF/ML measured less than that which FD-620 measured, and does the former take shorter time to measure?
- Why moisture content rate MX/MF measured is less than what FD-620 measured and does the former take shorter time to measure?
- What is minimum weight moisture content rate measured at?