Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the difference between MS/MX/MF/ML and Karl Fischer type analyzer?

analyzer karl fischer MF ML MS MX type
0
Posted

What is the difference between MS/MX/MF/ML and Karl Fischer type analyzer?

0

1. MS/MX/MF/ML is a heating and drying method analyzer that compares weight before and after heating and drying, while a Karl Fischer type analyzer titrates KF reagent that contains iodine to sample electric-chemically. 2. Karl Fischer method enables a measurement from some ppm to 100% (water), but operation is complicated and the unit is expensive. MS/MX/MF/ML is very easy to handle, needs short time to measure and is reasonably priced. 3. Where required resolution is under 0.01% MS/MX is more suitable in terms of handling, accuracy and running cost. There is no difference between data obtained with the Karl Fischer method and heating and drying method; however MS/MX is more likely to have better repeatability than a Karl Fischer type.