Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What is the maximum operating bandwidth of the high-voltage amplifier?

0
Posted

What is the maximum operating bandwidth of the high-voltage amplifier?

0

The load capacitance and output voltage together determine the maximum operating bandwidth of the Model 3211 high-voltage amplifier. For capacitive loads of less than 100 pF, the Model 3211 can output the full ±200 volts up to a maximum frequency of 600 kHz. For capacitive loads greater than 100 pF, the maximum bandwidth decreases. For 100-nF, the maximum bandwidth is about 800 Hz. For a 1000-pF load, the maximum operating bandwidth is approximately 80 kHz.