What is the reproducibility on the eye?
Reproducibility is better than 50 m. This corresponds to a mean postoperative error of less than 0.05D. In clinical use, mean standard deviation of 31,9 m was obtained for 15 measurements (Dr. Haigis, Univ.-Augenklinik Wuerzburg/Germany) and 33,4 m was obtained for 20 measurements (Dr. Vogel, University Eye Clinic, Mainz/Germany). The variability between 4 different examiners was 37,7 m (Dr. Haigis, Univ.-Augenklinik Wuerzburg/Germany) and between 5 different examiners was 29.8 m (Dr. Vogel, University Eye Clinic, Mainz/Germany). Safety criterion: If the results of the individual measurements differ by more than 100 m, no mean value will be displayed, and the measurement must be repeated.