What is the smallest quantity of DNA that can reasonably be measured with the BioPhotometer?

0
Posted

What is the smallest quantity of DNA that can reasonably be measured with the BioPhotometer?

0

The BioPhotometer should not be used to measure an absorbance of less than 0.05 at 260 nm. This OD value corresponds to a dsDNA concentration of app. 2.5 µg/ml (= app. 125 ng dsDNA in a 50 µl preparation). Please note that you should measure at 260 nm for a minimum absorbance of 0.1 in order to obtain exact results.