Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What speed and time should the BD Vacutainer® Blood Collection tubes be spun at?

0
Posted

What speed and time should the BD Vacutainer® Blood Collection tubes be spun at?

0

All non-gel blood collection tubes, including those that contain heparin, EDTA and non-gel serum tubes can be centrifuged at ≤1300 RCF for 10 minutes. The BD Vacutainer® SST™ and PST™ gel tubes should be spun at room temperature at a speed of 1000 to 1300 RCF for 10 minutes in a swinging bucket centrifuge and 15 minutes in a fixed-angle centrifuge.