Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What syringe size is recommended for use with Bard* catheters?

0
10 Posted

What syringe size is recommended for use with Bard* catheters?

0
10

It is recommended that a 10 cc syringe or larger be used to infuse or flush catheters. This includes pediatric and neonatal catheters. Infusion pressures should never exceed 25 psi because pressures higher than that may damage blood vessels and viscus. A two-pound weight equivalent force on the barrel of a 3 cc syringe generates pressure in excess of 25 psi. The same two-pound weight equivalent force on the barrel of a 10 cc syringe generates less then 8 psi of pressure. Please refer to the Instructions for Use (IFU) (IFU) for further details.