Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

What Transducers are most appropriate for measuring blood vessel diameter in-vivo?

0
Posted

What Transducers are most appropriate for measuring blood vessel diameter in-vivo?

0

The VD 5-2 sonomicrometer transducer pair can be sutured (order with patch) or attached with cyanoacrylate (order with no patch) across vessels greater than 2 mm in diameter. Alternatively, the Figure 1 low compliance spring-loaded clip can be used, primarily in acute preparations.