Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

When do you recommend using the 2-mm light path of the UVette and when should the 10-mm light path be used?

light path recommend Used uvette
0
Posted

When do you recommend using the 2-mm light path of the UVette and when should the 10-mm light path be used?

0

In general, the absorbance should be within the linear measuring range of the spectrophotometer (measuring range of Biophotometer: A = 0–3.0). For samples with a normal or a low concentration, the 10-mm light path is more suitable. For high-concentration samples (with A > 2.5), we recommend using the 2-mm light path. (An absorbance > 2.5 correlates to > 125 µg/ml dsDNA and > 100 µg/ml RNA respectively.