Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

When I set-up a high resolution PR balance the display comes up with ——– and cannot find zero. Why?

0
Posted

When I set-up a high resolution PR balance the display comes up with ——– and cannot find zero. Why?

0

The balance has not reached the operating temperature and is therefore unstable. It can take up to 15 minutes until a PR is warmed up and finds the zero point. Why does my AG/PG balance always ask for a calibration? The VariCal feature will ask for the balance to be calibrated once every twenty-four hours or if the weighing system has a 2 degree temperature change. VariCal can be turned off in the menu. Just turn INFO OFF in the menu. My balance shows flashing zeros (0). What can be done? Check to see if something is hitting the weighing pan. Make sure the balance is turned on with only the weighing pan and pan support on the balance. Why is my AE balance unstable? Check the calibration lever on the right-hand side of the balance. The lever should be toward the front of the balance. If it is not, pull it forward as far as it will go. Turn the balance off, then turn it back on. Calibrate the balance. On my HR73 or HG53 I keep getting ERROR3? The pan is on top of the panhandler and not i