Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

When quantifying a fluorescein labelled oligonucleotide by measuring absorbance at 260nm, what effect does the fluorescein have on this measurement?

0
Posted

When quantifying a fluorescein labelled oligonucleotide by measuring absorbance at 260nm, what effect does the fluorescein have on this measurement?

0
10

RESPONSE:The extinction coefficient of DNA at 260nm is around 10,000/mole/base or 10/mole/base. We use fluorescein isothiocyanate (Isomer 1) in the preparation of our fluorescein phosphoramidite (10-1963). The extinction coefficient of fluorescein (measured as FITC) at 260nm is around 13,700/mole or 13.7/mole. The fluorescein contribution to the absorbance of a fluorescein labelled oligonucleotide at 260nm is, therefore, about the same as 1 base or about 5% of a 20mer. This may be neglected or corrected for in the determination of the amount of labelled oligonucleotide from an A260 measurement. Note: dA=15.4, dC=7.4, dG=11.5, T=8.7 Fl-labeled ligos: 480nm excitation, 520nm emmission. The extinction coefficient for fluorescein-5-isothiocyanate at 495 nm is 76,000 L/mole-cm at pH 9. Upon conjugation to protein, and by analogy oligos, the extinction coefficient is decreased by 10%. This would result in a extinction coefficient of approximately 68,000 L/mole-cm. Additionally the extinction

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.