Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why can the width level of the WINDOW trigger on the DL1500 Series be set below the voltage sensitivity(V/div value)?

0
Posted

Why can the width level of the WINDOW trigger on the DL1500 Series be set below the voltage sensitivity(V/div value)?

0

The width level of the WINDOW trigger cannot be set to a value less than the voltage corresponding to 1 division at the specified voltage sensitivity. For example, if the voltage sensitivity is set to 500 V/div, then the width level cannot be set below 500 V. This limitation is due to the fact that the trigger sensitivity of the DL1500 Series is 1 divp-p.