Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Why won the PORTACOUNT allow fit testing if the ambient particle count is below 1000 particles per cubic centimeter?

0
Posted

Why won the PORTACOUNT allow fit testing if the ambient particle count is below 1000 particles per cubic centimeter?

0

If the ambient particle concentration is below 1000 particles per cm³ the number of particles that leak into the mask may not be sufficient to make an accurate measurement. An ambient particle concentration that low is rare. Usually it indicates that the PORTACOUNT is running low on alcohol.