Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Whats the difference between Vector/2 and EnviroLink?

envirolink vector
0
Posted

Whats the difference between Vector/2 and EnviroLink?

0

From an acquisition standpoint, Vector/2 and EnviroLink are nearly identical. Instrument tuning, calibration and data acquisition are essentially the same. A few instrument control parameters are hidden in the EnviroLink system aiding good laboratory practices for testing laboratories. Vector/2 has its own native processing, library and quantitation packages. EnviroLink, as a derivative of Vector/2, was introduced to take advantage of the HP ChemStation quantitation package. EnviroLink was introduced to target the environmental testing market. Currently, EnviroLink and Vector/2 normally ship with copies of HP (Agilent) ChemStation unless the purchaser specifies otherwise.