Whats the estimated bandwidth for the Model 162X nanosecond detector?
We do not specifically specify the bandwidth of the 162X. You can estimate, however, that on the 50-W setting the bandwidth is approximately 400 MHz. On the 10-kW setting, the bandwidth is approximately 1 MHz or greater. On the “Open” setting, the customer supplies an external load resistor. Since the bandwidth of the 162X depends on the size of the load resistor, we cannot specify the bandwidth in this mode.
Related Questions
- Can I use a 1-MW input on the oscilloscope with the Model 162X nanosecond detector and the Model 203X large-area receiver?
- We have poor internet bandwidth penetration in our premises. Can we use cloud computing S-a-a-S model in that case?