Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Are the criteria for testing stemmed femoral components in total hip arthroplasty still valid?

0
Posted

Are the criteria for testing stemmed femoral components in total hip arthroplasty still valid?

0

The increasing incidence of obesity in the population is a topic of current interest. This trend is reflected in patients undergoing primary Charnley hip replacement. Over a 21 year period from January 1986 to October 2006, during which 6910 primary operations were performed, the body mass of patients increased at a mean rate of 0.48 kg/year. The findings suggest that the test criteria for the endurance of stemmed femoral components be modified. The maximum load should be increased from 2.3 kN to 2.5 kN to reflect patients’ increasing weight. The duration of the test should be increased from 5 x 10(6) cycles during which the ‘femoral component shall not fracture’ to 10 x 10(6) cycles in order to reflect not only the patients’ activity level but also the 4-11 year period at risk for stem fracture, and so ‘reflect developing clinical experience’.