Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can I use a 935 controller with any detector and still obtain realtime collection and analysis?

0
10 Posted

Can I use a 935 controller with any detector and still obtain realtime collection and analysis?

0
10

No. The Model 935 is designed for use with a particular type of NaI detector that incorporates a highly efficient, proprietary high voltage bias supply built into the base of the detector. The Model 935 control unit provides 12 V dc power to the detector base along with a 0 – 1 Volt reference that controls the bias voltage for detector operation. The signal does, however, come back from the detector to the controller preamplifier/amplifier much like other systems. This system allows for fully automatic calibration where the Model 935 resets the bias voltage and amplifier gain settings.