Could C- and D-network mobile phones endanger patients with pacemakers?
OBJECTIVE: To investigate prospectively the extent of potentially harmful interference of cardiac pacemakers by mobile phones in the C (analog) and D (digital) networks in use in Germany. PATIENTS AND METHODS: 104 patients (54 men, 50 women; mean age 75.8 [40-100] years) with 58 different implanted pacemaker models (43 one-chamber and 15 two-chamber systems) underwent uniform tests at various functional states with three different telephones (D1 portable 8 Watt, D1 Handy model 2 Watt, C Handy model 0.5 Watt). The distances between telephone aerial and pacemaker, as well as reception sensitivity and polarity of the pacemaker were varied. All tests were done during continuous ECG monitoring. RESULTS: 28 different pacemaker types (48.3%) in 43 patients (41.3%) showed interference in the form of pacemaker inhibition and switching to interference frequencies as well as triggering of pacemaker-mediated tachycardias in the DDD mode, as well as in the temperature-regulated frequency-adaptive f