Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Do overshoots and undershoots of the block temperature occur with the Mastercycler® for the “block control” setting too?

0
Posted

Do overshoots and undershoots of the block temperature occur with the Mastercycler® for the “block control” setting too?

0

No. Only for the “tube control” setting. The block is then heated to above the programmed temperature (or cooled to below the programmed temperature) in order to reach the required temperature for the sample faster.