Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How did Tektronix find out about the problem?

problem tektronix
0
Posted

How did Tektronix find out about the problem?

0

Tektronix discovered the potential hazard after a customer instrument with a damaged power supply circuit was received for repair. In that particular case, the user had caused the defect by inserting a longer than specified screw or other thin metal object into the grounding post. Following a diligent analysis by Tektronix safety engineers, it was determined that this form of improper usage could have exposed the user to line-level voltage if the instrument had not been properly grounded.