How do I calibrate the Pressure Transducer for Esophageal Pressure Recording?
Data required for esophageal pressure recording can be obtained when the system is calibrated with a water manometer. The 0580Pes device is already calibrated when it leaves the factory. If you wish to perform additional calibration proceed as follows: Material required: one piece of ΒΌ” diameter (6.25mm) clear flexible tubing) six feet in length; two 3-way stop-cocks; two 10cc luer connector syringes; one piece of board for mounting the tubing; luer connectors as needed.