Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How do I calibrate the Pressure Transducer for Esophageal Pressure Recording?

0
Posted

How do I calibrate the Pressure Transducer for Esophageal Pressure Recording?

0

Data required for esophageal pressure recording can be obtained when the system is calibrated with a water manometer. The 0580Pes device is already calibrated when it leaves the factory. If you wish to perform additional calibration proceed as follows: Material required: one piece of ΒΌ” diameter (6.25mm) clear flexible tubing) six feet in length; two 3-way stop-cocks; two 10cc luer connector syringes; one piece of board for mounting the tubing; luer connectors as needed.