Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

How many analytes can be determined simultaneously using FS 3100?

analytes FS simultaneously
0
Posted

How many analytes can be determined simultaneously using FS 3100?

0

The FS 3100 can be configured with up to 2 FIA or 3 SFA channels. In principle, these can be run simultaneously. However, there are practical limitations in any lab environment, mostly relating to compatibility of sample matrices. For example, TKN and TP cannot be run simultaneously from the same sample cup as ammonia and ortho-phosphate (since they reside in an acidic digestion matrix incompatible with an aqueous matrix). Two, three, or four analytical channels are most often selected for simultaneous determinations. For example, in a wastewater treatment plant, ammonia, ortho-phosphate, nitrate/nitrite, and nitrite are often run as an analytical suite.