Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Is an edge target or a slit target more appropriate for MTF (Modulation Transfer Function) measurement?

0
Posted

Is an edge target or a slit target more appropriate for MTF (Modulation Transfer Function) measurement?

0

SBIR has chosen the edge target for use in its IRWindows standard target set. The signal from scanning across the edge target is differentiated to get the line spread function which would have been provided by a slit target. A Fourier Transform is then performed on this line spread function to generate the MTF. There are several advantages to the edge target over the slit target. The first is its general applicability. A slit target must be matched to the element size of the IR sensor to provide an accurate spread function. The slit width should be no more that 0.1 IFOV, with a narrower width even better. The target must be carefully manufactured to insure constant width over the height of the slit. A slit this small severely limits the flux passing through it, decreasing the signal-to-noise ratio. A separate target must be manufactured for each element size of interest. The edge target avoids all of these problems. A single edge can be used for any size element, and the manufacturing

Related Questions

What is your question?

*Sadly, we had to bring back ads too. Hopefully more targeted.

Experts123