Important Notice: Our web hosting provider recently started charging us for additional visits, which was unexpected. In response, we're seeking donations. Depending on the situation, we may explore different monetization options for our Community and Expert Contributors. It's crucial to provide more returns for their expertise and offer more Expert Validated Answers or AI Validated Answers. Learn more about our hosting issue here.

Can I test OAEs on patients with pressure equalization (PE) tubes?

0
Posted

Can I test OAEs on patients with pressure equalization (PE) tubes?

0

Yes. Patients with PE tubes can be tested by bypassing the auto-start function of the unit. To disable the auto-start, at the main menu select the ear to be tested by holding down the right or left arrow keys for 3 seconds until the green “test” light turns off. Once the key is released, the OAE unit will initiate testing.. If the auto-start function is not disabled by holding down the right/left arrow, the unit will get stuck in the auto-start phase and not initiate testing.